INESA Physico-Optical Instrument

INESA Physico-Optical Instrument Develops and produces globally trusted advanced physico-optical instruments, birthplace of China’s first polarimeter.

We are pleased to officially announce our participation in   LAB INDIA 2026, India’s leading trade fair for   instrument...
03/09/2026

We are pleased to officially announce our participation in LAB INDIA 2026, India’s leading trade fair for instrumentation, technology and .

📅 Date: September 10 – 12, 2026
📍 Booth: H5-D34
🏙️ Venue: HITEX Exhibition Centre, Hyderabad, India

We produce precision optical measuring , including , , -meters, -point-apparatus, , and more, designed to deliver accurate, repeatable results for quality control and &D laboratories across , , and industries.

🤝 Looking forward to meeting you and exploring new collaborations together!
🔗 Learn more about INESA Physico-Optical Instrument: www.shydwg.com

𝐌𝐞𝐥𝐭𝐢𝐧𝐠 𝐏𝐨𝐢𝐧𝐭 𝐀𝐩𝐩𝐚𝐫𝐚𝐭𝐮𝐬 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬🧪 Excerpted from our official  -point Apparatus knowledge handbook for lab QC and R&...
01/09/2026

𝐌𝐞𝐥𝐭𝐢𝐧𝐠 𝐏𝐨𝐢𝐧𝐭 𝐀𝐩𝐩𝐚𝐫𝐚𝐭𝐮𝐬 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬

🧪 Excerpted from our official -point Apparatus knowledge handbook for lab QC and R&D professionals.

Q1:Why does the -point-apparatus fail to output test results for a long time?
A1:
(1) The sample turns dark or carbonizes after melting. The WRS series melting point apparatus judges the initial and final melting points via optical signal changes. Therefore, the sample must be opaque before melting and transparent upon melting.
(2) Clogged furnace core causes low optical signal. Remove the furnace core per the manual, clean the light‑pass hole to ensure unobstructed light path, then retest.

Q2:Does the require after long‑term use?
A2: Annual calibration is recommended. You may perform self‑calibration using standard melting‑point substances (e.g. pure urea, p‑aminobenzoic acid), or send the instrument to a metrology institute for official verification.

Q3: Why is the sample melting range excessively long?
A3: A broad melting range usually indicates low sample purity; impurities disrupt the crystal lattice. It may also occur if the sample is a eutectic mixture, or undergoes decomposition during heating.

Q4: What is the impact of sample packing height on test results?
A4: Recommended packing height is 3 mm. Excessive packing height (>4 mm) slows internal heat transfer and creates large temperature gradients, resulting in prolonged melting range and higher measured values. Insufficient packing makes observation difficult.

Q5: Does sample particle size affect ?
A5: Yes. ‑particle samples have high thermal resistance, leading to lower measured melting temperatures. Therefore, samples must be fully ground before testing.

Q6: What happens if the sample is not packed tightly?
A6: Loose packing creates voids and poor heat conduction, broadening the melting range and deteriorating test reproducibility.

Q7: How to verify the of the instrument’s temperature sensor?
A7: Test using national standard reference materials with known precise melting points. If the deviation between measured value and standard value exceeds the allowable tolerance, calibration or repair is required.

&D

𝐏𝐫𝐞𝐜𝐢𝐬𝐢𝐨𝐧 𝐇𝐚𝐳𝐞 & 𝐓𝐫𝐚𝐧𝐬𝐦𝐢𝐭𝐭𝐚𝐧𝐜𝐞 𝐓𝐞𝐬𝐭𝐞𝐫 𝐟𝐨𝐫 𝐋𝐂𝐃 𝐎𝐩𝐭𝐢𝐜𝐚𝐥 𝐅𝐢𝐥𝐦𝐬Tired of unreliable test results on your LCD optical films?Ha...
27/08/2026

𝐏𝐫𝐞𝐜𝐢𝐬𝐢𝐨𝐧 𝐇𝐚𝐳𝐞 & 𝐓𝐫𝐚𝐧𝐬𝐦𝐢𝐭𝐭𝐚𝐧𝐜𝐞 𝐓𝐞𝐬𝐭𝐞𝐫 𝐟𝐨𝐫 𝐋𝐂𝐃 𝐎𝐩𝐭𝐢𝐜𝐚𝐥 𝐅𝐢𝐥𝐦𝐬

Tired of unreliable test results on your LCD optical films?
Hazemeter directly impact screen brightness, color quality and production yield — but thin, flexible films often cause distorted data and unstable QC.

𝐎𝐮𝐫 𝐒𝐆𝐖-𝟖𝟐𝟎 𝐬𝐨𝐥𝐯𝐞𝐬 𝐭𝐡𝐞𝐬𝐞 𝐩𝐚𝐢𝐧 𝐩𝐨𝐢𝐧𝐭𝐬:
✅ Integrating sphere optics for accurate, repeatable measurements
✅ Supports light sources A, C & D65 — freely select any 2 to fit your test protocols
✅ Dedicated magnetic clamp prevents sample deformation
✅ No darkroom needed, works seamlessly in labs and production floors
✅ Built-in data storage for full quality traceability

Verified on 0.1mm LCD-grade PET film (average):
- Transmittance: 90.71% (deviation: 0.03)
- Haze: 2.07% (deviation: 0.01)
High stability for enterprise-level batch inspection.

Compatible with all core LCD films: diffuser, BEF, reflective, polarizer, compensation film.
Compliant with ASTM D1003, ISO 14782, ISO 13468 and IEC TR 62977-2-5.

🙋 What’s your biggest headache with optical film testing? Drop a comment below.
📧 DM us for full product specs or overseas distribution opportunities.

𝐌𝐞𝐥𝐭𝐢𝐧𝐠-𝐏𝐨𝐢𝐧𝐭 𝐀𝐩𝐩𝐚𝐫𝐚𝐭𝐮𝐬 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬🧪Excerpted from our official  -point Apparatus knowledge handbook for lab QC and R&D...
25/08/2026

𝐌𝐞𝐥𝐭𝐢𝐧𝐠-𝐏𝐨𝐢𝐧𝐭 𝐀𝐩𝐩𝐚𝐫𝐚𝐭𝐮𝐬 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬

🧪Excerpted from our official -point Apparatus knowledge handbook for lab QC and R&D professionals.

Q1: How to properly pack samples into a capillary tube?
A1: The test sample shall be ground into fine powder. Unless otherwise specified, it shall be dried under the conditions stated in the corresponding drug monograph. When loading powder into the capillary, allow the powder to fall freely through a glass guide tube to ensure tight, uniform packing at the tube bottom. The recommended packing height is 3 mm to ensure consistent and compact sample loading.

Q2: If a capillary tube fractures inside the furnace core of the -68X Series Melting Point Apparatus, how should it be cleaned?
A2: ① Cut off the power supply and wait until the furnace cools down completely before operation; ② Open the heating furnace top cover, loosen the protective cover screws and pull out the guide post; ③ Hold the furnace core and pull out the capillary holder to remove all glass debris; ④ Align the notch of the capillary holder with the fl**ge for reassembly, restore all accessories, and verify smooth capillary insertion.

Q3: What reference materials are recommended for initial or when calibration data is abnormal? What are their final melting temperatures?
A3: Naphthalene (final melting point 80.6°C), Adipic acid (final melting point 153.0°C), Anthraquinone (final melting point 285.9°C).

Q4: What may cause inaccurate temperature readings on the melting point apparatus?
A4: Misaligned temperature sensor; contaminated heat-transfer liquid; motor failure that disrupts normal circulation of the heat-transfer liquid.

Q5: What causes heating failure on a microthermal melting point apparatus?
A5: ① Loose connecting cable between the heating stage and program control unit → Inspect the cable and ensure secure, reliable plug connection; ② Broken heating wire → Confirm the mains power supply is normal and the instrument power switch is turned on; ③ Abnormal power supply or control circuit board failure → Do not disassemble the instrument without authorization. Contact the original manufacturer for professional repair directly.

Q6: What leads to abnormal temperature display on the melting point apparatus?
A6: Damaged platinum resistance sensor or analog circuit board. The issue can be resolved by replacing the platinum resistance sensor and performing full temperature recalibration.

Tired of switching between separate instruments for wet and dry particle sizing, or dealing with inconsistent results ac...
20/08/2026

Tired of switching between separate instruments for wet and dry particle sizing, or dealing with inconsistent results across sample types?

🏅 Our laser particle size analyzer handles it all in one system. Built on full-range Mie scattering theory, it delivers 0.02–2000μm (wet) / 0.1–2000μm (dry) measuring ranges with ±0.5% repeatability, compliant with and .

One-click mode switching, results in quick time — ideal for pharmaceuticals, battery materials, fine chemicals and food powders.

📧 Drop "1" for the brochure | ☎️Inquire NOW: [email protected]

𝐌𝐞𝐥𝐭𝐢𝐧𝐠-𝐏𝐨𝐢𝐧𝐭 𝐀𝐩𝐩𝐚𝐫𝐚𝐭𝐮𝐬 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬Excerpted from our official  -point Apparatus knowledge handbook for lab QC and R&D ...
18/08/2026

𝐌𝐞𝐥𝐭𝐢𝐧𝐠-𝐏𝐨𝐢𝐧𝐭 𝐀𝐩𝐩𝐚𝐫𝐚𝐭𝐮𝐬 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬
Excerpted from our official -point Apparatus knowledge handbook for lab QC and R&D professionals.

Q: What is the core working principle of Physico-Optical instruments' melting‑point apparatus for substance melting‑point testing?
A: -substances reflect light, while -substances transmit light. An abrupt change in light transmittance occurs during the melting process. The instrument detects this change automatically via sensing to identify the incipient‑melting and complete‑melting temperatures.

Q: What melting‑point determination methods are specified in the Chinese , and which test samples do they apply to?
A: There are 3 methods:
Method 1 applies to easily solid substances.
Method 2 applies to ‑crushable solid samples (e.g. , , ).
Method 3 applies to petrolatum and other similar substances.

Q: What are the specification requirements for melting‑point in Method 1?
A: The capillary is a hard‑glass tube with an inner diameter of 0.9–1.1 mm, wall thickness of 0.1–0.15 mm, and one end sealed. Its length shall allow immersion in the ‑transfer liquid so that the sample content sits at the middle of the thermometer bulb.

Q: How does -rate affect melting‑point measurement?
A: An excessively fast heating rate leads to higher measured temperature and shorter melting range.
An excessively slow heating rate may cause softening, decomposition or sublimation of the sample, resulting in lower measured temperature.
A slow heating rate (e.g. 1 °C/min) is recommended for purity testing for precise results.
The starting temperature is usually set 3–15 °C below the expected melting point of the sample and adjusted according to the heating rate.

Q: What methods can avoid for samples that decompose while melting?
A: Adjust the heating rate to 2.5–3 °C/min.
Take the temperature at which the sample starts partial liquefaction or bubble generation as the incipient‑melting temperature.
Take the temperature of full sample liquefaction as the complete‑melting temperature.
If solid‑phase disappearance is indistinct, use the temperature at which the sample starts expanding and rising as the complete‑melting temperature.
For samples where incipient and complete melting cannot be told apart, take the temperature of abrupt physical change as the melting point.

🧪 Tired of subjective visual color assessments creating batch-to-batch inconsistencies and quality disputes?Our  ‑2 Tint...
13/08/2026

🧪 Tired of subjective visual color assessments creating batch-to-batch inconsistencies and quality disputes?

Our ‑2 Tintometer eliminates subjective judgment bias by adopting the globally‑recognized visual color‑grading scale for liquid sample assessment. Its dual‑field optical design enables direct side‑by‑side comparison between samples and calibrated standard color filters, delivering consistent color grading for edible oils, fats, chemical solutions, cosmetics and more.

No complex setup or calibration required, unify your QC standards across incoming inspection, production and finished product testing with traceable, industry-aligned results.

✉ Drop "1" for the brochure | ☎️Inquire NOW: [email protected]

🧪 𝐏𝐨𝐥𝐚𝐫𝐢𝐦𝐞𝐭𝐞𝐫 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬Excerpted from our official   knowledge handbook for   and  &D professionals.1️⃣Q: A temperatu...
11/08/2026

🧪 𝐏𝐨𝐥𝐚𝐫𝐢𝐦𝐞𝐭𝐞𝐫 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬

Excerpted from our official knowledge handbook for and &D professionals.

1️⃣Q: A temperature-controlled polarimeter activates the temperature control function but fails to regulate the temperature — what are the common causes and solutions?

A: ① The temperature probe is not inserted into the sample tube's temperature port → Insert the probe correctly and secure it tightly.

② No temperature-controlled tube is loaded, or temperature control is activated with an empty chamber → Remove the tube and disable temperature control first.

③ The temperature control board has no ±10 V output, or the Peltier element is damaged → Replace the temperature control board / semiconductor cooling chip.

2️⃣Q: What advantages does the offer over the ?

A: The runs on the Android OS with superior scalability, enabling role-based user management and structured data management for seamless integration with your . An optional software module fully compliant with 21 CFR Part 11 is also available.

3️⃣Q: The temperature-controlled tube reads normally, yet heating/cooling never stops — what is the most common operator-related cause?

A: The operator removes the temperature-controlled tube without first turning off the temperature control, leaving the probe suspended in air and continuously responding to ambient temperature.

4️⃣Q: How does the Chinese (ChP), Volume IV, Section 0621 — Optical Rotation Determination — define positive and negative rotation?

A: Substances that rotate the plane of polarized light to the right (clockwise) are classified as dextrorotatory and denoted with a "+" symbol; those that rotate it to the left (counterclockwise) are levorotatory and denoted with a "−" symbol.

5️⃣Q: What does daily look like?

A: Store the instrument in a well-ventilated, dry place to prevent moisture damage. Empty the solution from the sample tube immediately after use and rinse thoroughly with distilled water. Wipe optical lenses only with a soft lint-free cloth — never touch them with bare fingers.

🧪Need reliable turbidity testing on site, not just in the lab?Our  -B series portable turbidimeters come in 4 models to ...
06/08/2026

🧪Need reliable turbidity testing on site, not just in the lab?

Our -B series portable turbidimeters come in 4 models to match every application: WGZ-20B (0–20 NTU), WGZ-200B (0–200 NTU), WGZ-1000B (0–1000 NTU) and WGZ-2000B (0–2000 NTU).

Built on ratio measurement technology with 860nm and fully compliant with , all models deliver stable readings even for colored samples, with automatic range switching, multi-unit support (NTU/FTU/EBC/ASBC) and 2000-set data logging.
From ultrapure drinking water to high-turbidity industrial wastewater, one product family handles all your field QC needs.

✉ Drop "1" for the brochure | ☎️Inquire NOW: [email protected]

🧪𝐏𝐨𝐥𝐚𝐫𝐢𝐦𝐞𝐭𝐞𝐫 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬Excerpted from our official polarimeter knowledge handbook for lab QC and R&D professionals.Q: ...
04/08/2026

🧪𝐏𝐨𝐥𝐚𝐫𝐢𝐦𝐞𝐭𝐞𝐫 𝐏𝐫𝐨𝐝𝐮𝐜𝐭 𝐅𝐀𝐐𝐬
Excerpted from our official polarimeter knowledge handbook for lab QC and R&D professionals.

Q: When testing an optical rotation sample, the measured reading is slightly higher than the standard value and results in non-compliance. What are the potential causes?
A: Tare weighing was not performed during sample weighing; the protective sheet and washer of the sample tube were installed backwards; the tube’s end caps were overly tightened; deviation existed in blank measurement.

Q: What are the differences in minimum readout and indication error between model SGW-568A and SGW-533A?
A: SGW-568A: minimum readout 0.0001°, relative error ±0.003°; SGW-533A: minimum readout 0.001°, relative error ±0.004°. Both models have an indication error of ±0.01° at 589 nm.

Q: What solutions can be adopted for dark-colored samples with poor light transmittance that cannot be measured?
A: Dilute the sample, use a shorter sample tube, or switch to a model with high light transmittance.

Q: What requirements should be followed when filling samples into the tube?
A: Move air bubbles in the bulb-type glass tube to the bulb neck; wipe off water droplets on the light-transmitting surfaces with a soft cloth; do not over-tighten the screw cap to avoid mechanical stress that distorts readings; align the marked position and direction each time the tube is inserted.

Q: What core structural upgrades does SGW-531/532 have compared with SGW-1/2, and what operational advantages do these upgrades bring?
A: High-precision encoders replace the original worm gear and worm drive. This eliminates mechanical noise, speeds up measurement, and improves measuring accuracy.

Q: A customer gets unstable readings when testing optical rotation samples. What are the possible reasons?
A: Air bubbles in the sample liquid, light source attenuation, encoder malfunction, oxidation of the tube holder, and over-tightened end caps on both sides of the sample tube.

Q: What are the typical service lives of LED light sources, tungsten halogen lamps and sodium lamps?
A: LED light sources last approximately 100,000 hours. Tungsten halogen lamps and sodium lamps have far shorter lifespans: sodium lamps work for several thousand hours, while tungsten halogen lamps last several hundred to several thousand hours.

Address

No. 88 Xutang Road , Songjiang District Shanghai
Shanghai
201613

Opening Hours

Monday 08:30 - 16:00
Tuesday 08:30 - 16:00
Wednesday 08:30 - 16:00
Thursday 08:30 - 16:00
Friday 08:30 - 16:00

Alerts

Be the first to know and let us send you an email when INESA Physico-Optical Instrument posts news and promotions. Your email address will not be used for any other purpose, and you can unsubscribe at any time.

Shortcuts

Share