A sterilization cycle that passed qualification on the first study.
For Pharma Swede, EPAES delivered a Steam-In-Place (SIP) system end to end: site survey, engineering, procurement, installation and commissioning, through to performance qualification.
The system sterilizes the holding tank, process piping and sterilizing filters in place, in six controlled phases:
1. Preparation and isolation
2. Nitrogen purge
3. Air removal and heat-up
4. Sterilization hold at 121 °C or more
5. Cool-down with a sterile nitrogen vacuum break
6. Post-SIP hold and documentation
Performance qualification by an independent validation contractor:
• 12 temperature loggers and 1 pressure logger, with a biological indicator at every sensor
• 3 consecutive cycles: 3 out of 3 passed
• Every sensor held 121 °C or more
• Lowest F0 of 44.97 min at the coldest point, against a limit of 20
• Loggers calibrated before and after the study, largest deviation 0.19 °C
Project engineers: Eng. Mohamed Farouk and Eng. Ziyad Hamad